Syringe Pump Agilia SP MC
Flow rate range
0,1 – 1200 ml/h, depending on the syringe capacity (0,1 ml/h increment). 0,01 minimum increments instead of 0,1 ml/h is activated when the max patient weight is set to 20 kg in the Agilia Vigilant drug library. Flow rate can be limited according to the medicine name (soft and hard limits) with Agilia Vigilant Drug’Lib, IV Medication Safety System.
Flow rate accuracy
±1 % on mechanism : ± 2 % on syringes.
Syringes capacities
5, 10, 20, 30/35, 50/60 cc.
Type of syringe
Up to 100 types.
Infusion Modes
ml/h mode.
Dose rate mode: ng/h, ng/kg/min, ng/kg/h, microg/min, microg/h, microg/kg/min,microg/kg/h, mg/min, mg/h, mg/24h, mg/kg/min, mg/kg/h, mg/kg/24h,mg/m2/h, mg/m2/24h, g/h, g/kg/min, g/kg/h, g/kg/24h, mmol/h, mmol/kg/h,mmol/kg/24h, mU/min, mU/kg/min, mU/kg/h, U/min, U/h, U/kg/min, U/kg/h,kcal/h, kcal/24h, kcal/kg/h, mEq/min, mEq/h, mEq/kg/min, mEq/kg/h.
Dilution setting: — units / ml or — units / — ml.
With or without loading dose.
Volume or dose / time: 0,1 – 99,9 ml; 00 h 01 – 96 h 00.
Volume limit: 0,1 – 999 ml
Volume/Dose infused
Volume : 0,1 – 999,9 ml / Dose: 0,1 – 999,999 units
Priming
3 modes: mandatory, not mandatory, or advised / Rate: 1200 ml/h.
Bolus
– Direct bolus: Rate: 50 – 1200 ml/h (50 ml/h increment)
– Programmed bolus (dose or volume / time): 0,1 – 99,9 ml ; 0,1 – 9999 units / 1 min – 24 h.
Loading dose
Dose or volume / time: 0,1 – 99,9 units / 00 min 01 – 59 min 59 rate auto-calculation.