Agilia SP
Agilia SP MC
Agilia SP PCA
Agilia SP TIVA
Agilia VP
Agilia VP MC
Infusia VP7
Link + Agilia®
CONOX
MRI Guard Agilia®
Amika®
CATSmart
Agilia SP | Basic Pump
- Flow rate range: 0.1- 200 ml/h (depending on the syringe capacity – up to 1200 ml/h).
- Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.
- Syringe capacities: 5, 10, 20, 30/35, 50/60 CC.
- Infusion Modes:
- ml/h (0.1 ml/h increments).
- Volume / time: Volume to infuse 0.1 to 99.9 ml /
- Time: 0 h 01 to 96 h 00.
- Volume limit: 0.1-999.9 ml (0.1 ml increments).
- Bolus: Manual bolus: Rate 50-1200 ml/h (50 ml/h increments).
- Drug library: Initial drug list of 50 agents adjustable with Vigilant®, the IV Medication Safety Solution. Adjustment of the dilution, authorised infusion modes, default infusion modes, flow rate and bolus limits in ml/h.
- Pressure modes: 2 modes available: variable or 3 pre-set levels – From 100 to 900 mmHg, 50 mmHg increment.
- DPS: The Dynamic Pressure System – DPS – warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
- Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml max).
- Dimensions (H/W/D/) / Weight: 160 x 345 x 135 mm / 2150 g.
- Battery: NI/MH / Battery capacity minimum: 10 hours at 5 ml/h / Battery replacement: separate compartment – no need for electrical security testing.
- Power supply: 100 – 240 Vac / 15 VA, external power 9 Vdc by a converter.
Agilia SP TIVA
- Flow rate range: 0.1-1200 ml/h (depending on the syringe capacity).
- Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.
- Syringe capacities: 5,10, 20, 30/35, 50/60 CC.
- Infusion modes
- ml/h.
- Dose rate.
Dose units: ng, μg, mg, cal, kcal, U, kU, mmol, mol, Kmol / kg.
Time units: 24h / h / min.
Dilution setting: X units / ml or X units / Y ml. - Volume or dose / time: 0.1-99.9 units ; 00h01- 96h00.
- Volume limit: 0.1-999.9 ml.
- TCI Functions: Injectomat® TIVA Agilia is an IV anaesthesia system incorporating pharmacokinetics models for Propofol, Remifentanil, Sufentanil in plasma or effect mode.
- Bolus: Manual bolus: Rate: 50 -1200 ml/h; 50 ml/h increment. Programmed bolus (dose or volume/time): 0.1-99.9 units / 0.1-1200 ml/time auto-calculation.
- Programming modes: No drug names: infusion without any display of the drug names. Drug labelling: infusion with the drug names. Vigilant® Drug’Lib: the secured IV Administration mode with a drug library.
- Drug labelling: Display of the drug name based on a list of 50 pre-set drug names.
- Drug libraries: Drug parameters, infusion, bolus and loading dose parameters, authorised infusion modes, infusion mode per default. Inititial drug library of 50 drugs adjustable with Vigilant®, the IV Medication Safety Solution. Up to 120 drugs (depending on comment size) storable in 4 different drug libraries.
- Pressure modes: 2 modes available: variable or 3 pre-set levels – From 100 to 900 mmHg, 50 mmHg increment.
- DPS The Dynamic Pressure System – DPS – warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
- Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml max).
Agilia VP
- Flow rate range: 0.1-1200 ml/h. Flow rate can be limited according to drug name
- Flow rate accuracy: +/- 5% and even better in most clinical situations
- Syringe capacities: 5,10, 20, 30/35, 50/60 CC.
- Infusion modes
ml/h mode, Volume + Flow rate, Volume + time, Flow rate + time, Volume + time + Rate - Infusion OCS Patent:
The Occlusivity Check System is the only auto-test to check the good working order of the pump in association with its set and thus preventing risk of free-flow - Bolus: Direct bolus: Rate 50-1200 ml/h (50ml /h increment)
- Pressure modes:
- No drug name: Infusion without any display of the drug names
- Drug Labelling: Infusion with the drug names displayed
- Agilia Vigilant Drug Lib: the secured IV administration mode by soft and hard limit
- 2 Pressure Modes: Variable or 3 pre-set levels –Range from 50 – 750mmHg.
- DPS: The Dynamic Pressure System – DPS – warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated
-
Dimension (H/W/D) / Weight: 1135 X 190 X 170 mm / ~ 2 kg
- Battery: Li-ion Smart Battery, Battery life (when fully charged): >8 h at 125ml/h
- Power supply: 100V – 240V ~50/60Hz with functional earth
Agilia VP MC
- Flow rate range: 0.1-1500 ml/h, Flow rate can be limited according to drug name
- Flow rate accuracy: +/- 5% and even better in most clinical situations
- Infusion modes
- ml/h mode, Volume + Flow rate, Volume + time, Flow rate + time, Volume + time + Rate, Ramp-up / Ramp-down, Sequential / Intermittent, Secondary / Piggyback
- Dose rate modes: Dose units: ng/h, ng/kg/min, ng/kg/h, μg/min, μg/h, μg/kg/min, μg/kg/h, mg/min, mg/h, mg/24h, mg/kg/min, mg/kg/h, mg/kg/24h, mg/m²/h, mg/m²/24h, g/h, g/kg/min, g/kg/h, g/kg/24h, mmol/h, mmol/kg/h, mmol/kg/24h, mU/min, mU/kg/min, mU/kg/h, U/min, U/h, U/kg/min, U/kg/h, kcal/h, kcal/24h, kcal/kg/h, mEq/min, mEq/h, mEq/kg/min, mEq/kg/h, mL/kg/24h, mL/kg/h.
- Infusion OCS Patent:
The Occlusivity Check System is the only auto-test to check the good working order of the pump in association with its set and thus preventing risk of free-flow - Bolus: 1500 ml/h Adjustable from 50mL/h to 1500 mL /h by 50mL/h increments Direct or Programmable
- Programming Modes:
- No drug name: Infusion without any display of the drug names
- Drug Labelling: Infusion with the drug names displayed
- Agilia Vigilant Drug Lib: the secured IV administration mode by soft and hard limit
- 2 Pressure Modes: Variable or 3 pre-set levels –Range from 50 – 750mmHg.
- DPS: The Dynamic Pressure System – DPS – warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated
- Dimension (H/W/D) / Weight: 1135 X 190 X 170 mm / ~ 2 kg.
- Battery: Li-ion Smart Battery, Battery life (when fully charged): >8 h at 125ml/h
- Power supply: 100V – 240V ~50/60Hz with functional earth
Infusia VP7
- Flow rate range:
- 1.0 ml/h – 1200 ml/h
- 1.0 – 99.9 ml/h, 0.1 ml/h increment
- 100 – 1200 ml/h, 1 ml/h increment
- 1 – 400 drops/min, 1 drop/min increment
- Flow rate accuracy: ± 5 %
- Volume to infuse: 0 – 9999 ml, 1 ml increment
- Infusion modes
- Rate setting mode – ml/h, drops/min
- Time setting mode – minutes, ml/h, drops/min
- Volumetric setting mode – ml/h
- Time of infusion: 1 – 1999 min, 1 min increment minimum
- Priming: Manual or by using the purge/bolus function of the pump
- Bolus rate:
- 1 – 1200 ml/h
- 0.1 – 99.9 ml/h, 0.1 ml/h increment
- 100 – 1200 ml/h, 1 ml/h increment
- Purge rate: 600 ml/h
- Free flow protection: Anti-free flow clamp when door is closed
-
End of infusion: Automatic KVO at the end of infusion: 1 ml/h if < 10 ml/h infusion rate, otherwise, 3 ml/h
-
Data log event: 5,000 events
Link + Agilia®
Features
More Safety
-
- Central display of the IV pumps and Link+ Agilia status is a safety enhancement of the new user-centered design.
- Easy-to-clean design ensures better disinfection.
- Connected to your Nurse Call System, Link+ Agilia saves you time.
More convenience
-
- Link+ Agilia stacks, powers and centralizes data for 4, 6 or 8 Agilia IV Infusion pumps and reduces the number of power cables to one.
- IV pump installation and removal are easy on the Link+ Agilia.
More efficiency
-
- Captures and stores infusion data automatically in your hospital IT system.
- The central alarm display is visible 360º from a distance.
CONOX
Two monitoring solutions in one device, using one single sensor. CONOX is a non-invasive depth of anesthesia monitor designed to help the anesthesiologist assess patient brain activity and to rapidly detect how certain drugs are affecting the patient.
MRI Guard Agilia®
MRI Guard Agilia takes Agilia infusion pumps into magnetic resonance units.
The MRI Guard Agilia is shielded externally to prevent interference generated by the Agilia infusion pumps from adversely affecting the MR image quality, and it shields the infusion pumps against electromagnetic disturbances from the MRI system.
Amika®
Amika is an ambulatory volumetric pump intended to be used for administration of fluids through transnasal or percutaneous feeding tubes. Amika delivers fluids in a continuous mode using a linear peristaltic pumping mechanism.
Features
Designed for more safety
- Dedicated enteral giving sets designed with ENPlus and ENLock
- Latest sensor technology for enhanced detection capability
- Keypad and Settings Lock functions for a safer use of Amika® by the patient
Intuitive for higher efficiency
- Graphical symbol-based display for rapid and universal understanding
- Ergonomic and robust design suitable for hospital and homecare
- Direct access to main functions to optimize workflow
Convenient for better comfort
- Silent operating
- Compact design and long battery life for high mobility
- Multiple possibilities of pump settings to adapt to all environments
- Versatile pump holder usable with a wide range of supports
CATSmart
Fresenius Kabi introduces CATSmart. The unique continuous red cell separation process based on Continuous Flow technology is designed for recycling blood intra- and post-OR during cardiac, orthopedic or vascular surgery. CATSmart is the only autotransfusion device on the market using Continuous Flow technology which guarantees an early access to RBC at any time with a reliable quality.
Features
Clinical performance benefits:
-
- Consistently high hematocrit
- Excellent wash out performance of albumin, potassium, heparin, free hemoglobin
- Complete elimination of non-emulsified fat
- Fastest processing time due to Continuous Flow technology
Ergonomic benefits:
- Volume independent „one-fits-all” washing chamber
- Unique height adjustment for better handling
- Autostart function
- Intelligent and silent modular vacuum source
- A colored intuitive touchscreen
- Efficient data transfer management